The probes of the EP-600 series can be defined as the all-in-one that set the standard for miniature and optical communication in broadband sensors of the electric field.
Exceptionally small-sized spherical symmetrical configuration, lightweight and miniaturized electronics combined with excellent RF characteristics make the EP-600 series RF Electric Field Probes the ideal solution for all EMC/EMI applications (chambers and TEM/GTEM cells), biology and materials research and wherever fast and accurate measurements with negligible or minimum interference of the probe to the field under measure are essential.
Accredited calibration on request.
Probes EP-600 series are available in the following versions
Комплектация
Сведения о реестре СИ
Номер в Госреестре |
Наименование СИ |
Обозначение типа СИ |
Изготовитель |
Срок свидетельства или заводской номер |
---|---|---|---|---|
64859-16 |
Пробники электрического поля |
PMM EP-600, PMM EP-601, PMM EP-602, PMM EP-603 |
Фирма "Narda Safety Test Solutions S.r.l.", Италия |
31.08.2021 |
Номер в Госреестре |
Наименование СИ |
Обозначение типа СИ |
Изготовитель |
Срок свидетельства или заводской номер |
---|---|---|---|---|
64859-16 |
Пробники электрического поля |
PMM EP-600, PMM EP-601, PMM EP-602, PMM EP-603 |
Фирма "Narda Safety Test Solutions S.r.l.", Италия |
31.08.2021 |
COMMON SPECIFICATIONS |
|
Resolution | 0,01 V/m |
Sensors | Six monopoles |
Measurement mode | X-Y-Z axis simultaneous sampling |
Sampling rate | 22 to 0,03 samples/s depending on filter settings |
Digital filtering | Low-pass 2,3 to 28 Hz; 80 dB notch 50/60 Hz |
Internal battery | 3V, 5 mA/h, Li-Mn rechargeable |
Operation time | 100 hours @ 0,4 S/s, 28 Hz filter |
Recharging time | 48 hours for full operation time Partial charging allowed, e.g.: 8 hrs rech. = 12 hrs operation @ 5 S/s 1 hr rech. = 5 hrs operation @ 0,4 S/s |
Internal data memory | Serial no. - Calibration factors & date Firmware version |
Communication | Bidirectional fiber optic link |
Fiber optic connector | HFBR-0500 |
Fiber optic length | 10 m, standard 20/40 m options |
Fiber optic to PC | Fiber optic to RS232 converter RS232 to USB converter |
PC software | Displays field, temperature, battery, sampling, filters, averaging, frequency setting, manual data logging as text file |
Operating temperature | -10 °C + 50 °C |
Temperature reading resolution | 0,1 °C |
Battery voltage reading resolution | 10 mV |
Probe holder tip | 1/4 20 UNC female |
Main specifications (see the user manual for complete information) |
|
EP-600 | |
Frequency Range | 100 kHz to 9,25 GHz |
Flatness |
Frequency correction OFF 1 to 150 MHz: 0,8 dB 0,5 to 6000 MHz: 1,6 dB 0,3 to 7500 MHz: 3,2 dB Frequency correction ON, typical 0,3 to 7500 MHz: 0,4 dB |
Dynamic Range | 0,14 to 140 V/m 60 dB single range |
Overload | 300 V/m |
Dimensions (overall) | 53mm |
Weight (incl. 1 m fiber optic pigtail) | 23 g |